KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone +49-[0]7433- 9933-0 Fax +49-[0]7433-9933-149 Internet: www.kern-soh
3.4 Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testi
6 Unpacking, Setup and Commissioning 6.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are a
Ensure that transport guard has been removed (only existing in 6 kg models) To loosen the transport guard screw out transport screw [1] anticloc
6.2.2 Scope of delivery Serial accessories: • Balance • Weighing plate • Power cable • Protective cover • Instruction Manual 6.3 Mains con
6.7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance
Press the TARE button, the acceleration value due to gravity is displayed. Press the TARE key anew. „UnLoAd“, followed by „LoAd“ is display
6.7.2 Models CPB-DM (verified models) The adjustment is locked for verified balances. Carrying out adjustment requires that the seal is destroyed a
TARE button, „UnLoAD“ appears Press the TARE key anew „LoAd“ as well as the currently set adjustment weight appear flashing either • Confirm
6.8 Linearization (non-verified models only) Linearity shows the greatest deviation of a weight display on the scale to the value of the respective t
Press TARE button, the acceleration value due to gravity is displayed. Press the TARE key anew „LoAd 0“ appears, after a short time a sign
GB KERN CPB-N / CPB-DM Version 2.4 08/2014 Instruction Manual Counting balance Contents 1 Technical data ...
Second adjustment weight after a short time a signal will sound and „LoAd 0“ appears. First adjustment weight After successful linearization
6.9 Verification General introduction: According to EU directive 90/384/EEC balances must be verified if they are used as follows (legally control
6.9.1 Adjustment switch and seals Possible seals: enforced, and or 1. Seal mark 1 2. Cover 3. Verification switch 4. Verification wire
6.11 Service mode (verified models) This overview of the service parameters is merely for checking the parameters set by the appropriate Bureau of Sta
Overview for service parameter: * = Factory settings = The menu item is blocked; to modify press the justification button. Menu block Main menu Me
F9 Grv Gravity not documented F10 bEP Audio signal ok* Signal sound, when the load is within the set range Low Signal sound, when the load is
7 Operation Mode 7.1 Switch on/off and set zero Operation Indication 1. Switch on balance Press ON/OFF switch and hold briefly (at the bottom rig
7.3 Weighing with taring The dead weight of any weighing container may be tared away by pressing a button, so that the following weighings show the
Reset display to "0": The total weight of the container is tared away. Add more components into the weighing container (adding). Now
7.4 Display background illumination The balance has the following possibilities of setting for the display background illumination: Keep pressed
7 Operation Mode ... 27 7.1 Switch on/off and set zero ...
8 Piece counting With pieces counting you can either count parts into a container or remove parts from a container. To count a greater number of part
Count the items: Tare if necessary, place weighing good and read off the number of items. Weight placed on balance: Reference weight: Quantity plac
8.2 Numeric entering of the reference weight If you know the reference weight/piece you can enter this via numeric keys. Set reference: Enter refere
8.3 Automatic correction of the reference weight (verifiable models only) The balance has the possibility to correct the deviation automatically if
9 Totalization This function allows you to execute several weighing procedure. After that, the total items number, the total weight and the number of
Printout example KERN YKB-01N: Number weighing processes NO. 1 Weight placed on balance GS 2.000 kg Reference weight U.W. 100 g Quantity placed o
Add more weighed goods as described before. Please note that the weighing system must be unloaded between the individual weighing procedures.
9.2 Automatic adding-up With this function the individual weighing values are automatically added into the summation memory when the balance is unloa
10 Weighing to target quantity or target weight and tolerance check An acoustic signal is sounded as soon as the number of items placed or a certain w
Set limit values: Determine upper limit value for target item number: Press , „Hi Cnt“ will appear: Enter the upper limit value via the nume
1 Technical data KERN CPB 6K0.1N CPB 15K0.2N CPB 30K0.5N Readability (d) 0.0001 kg 0.0002 kg 0.0005 kg Weighing range (max) 6 kg 15 kg 30
Start tolerance check Determine the average piece weight (see chap. 8.1) or enter manually (see chap. 8. 2). Place the load, wait until the aco
Select target weight mode: By keeping pressed the desired tolerance weighing mode can be selected: • Check off – Tolerance check switched off • C
Determine lower limit value for target weight: Enter the lower limit value via the numeric keyboard, e.g.: 90 g Confirm within 5 sec: Press
12 Menu (non verifiable models) Access to menu: Switch-on balance and during the selftest press the TARE button. Use the number keys to ente
12.1 Menu overview Menu block Main menu Menu item Submenu Available settings / explanation F1 CAL Adjustment F2 di d 6000* Resolution d 3000
F9 Gru Gravity not documented F10 bEP Audio signal ok* Signal sound, when the load is within the set range Low Signal sound, when the load is b
13 Data output RS 232 C The balance is typically equipped with a RS 232C interface. Weighing data can be edited according to menu setting or automati
14 Service, maintenance, disposal 14.1 Cleaning Before cleaning, please disconnect the appliance from the operating voltage. Please do not use aggre
15 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must
16 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax: 0
KERN CPB 6K1DM CPB 15K2DM CPB 30K5DM Readability (d) 0.001 kg; 0.002 kg 0.002 kg;0.005 kg; 0.005 kg;0.01 kg; Weighing range (max) 3 kg; 6 kg 6
KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet
NL Wij verklaren hiermede op exclusieve verantwoordelijkheid dat het product, waarop deze verklaring betrekking heeft, met de hierna volgende richtli
2 Appliance overview 1. Weighing plate / rechargeable battery compartment (under weighing plate) 2. Bubble level 3. RS 232 interface 4. Foo
2.1 Overview of display Weight Reference weight Quantity Battery charge status display 2.1.1 Display weight Here the weight of the load is d
2.2 Keyboard overview Selection Function • Numeric keys • Deleting key • Call up target quantity and target weight mode • Addition in sum
3 Basic Information (General) 3.1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed. It i
Komentarze do niniejszej Instrukcji