Kern-sohn EMB 500-1GN Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Waga Kern-sohn EMB 500-1GN. KERN & SOHN EMB 500-1GN User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 0
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating Instructions
Precision balance
KERN EMB 500-1GN
Version 1.0
05/2014
GB
EMB-BA-e-1410
Przeglądanie stron 0
1 2 3 4 5 6 ... 15 16

Podsumowanie treści

Strona 1 - Precision balance

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-so

Strona 2 - KERN EMB 500-1GN

6 Operation 6.1 Overview of the keypad Key Pressed briefly Pressed for 10 sec Switching the balance on Tara-Function Unit changing Switc

Strona 3 - 1 Technical data

After removal of the tare cup the total weight appears as a negative value with „-„ mark. 6.2.3 Units Different foreign weighing units are integrat

Strona 4

 Press the ON/TAREkey to switch among the settings: Display Mode „tr“ ON Auto-Zero on „tr“ OFF Auto-Zero off  Confirm with . Wait until the b

Strona 5 - 4 Transport and storage

6.2.6 Reset function Factory setting reset.  When the balance is switched off, press and keep it pressed  Do not release . Press also the -ke

Strona 6

7 Maintenance, upkeep, disposal 7.1 Cleaning Please disconnect the device from the operating voltage before cleaning. Only use a cloth dampened wi

Strona 7 - x 9 V Block

8 Troubleshooting The balance should be switched off for a short time following an interruption in the programme sequence and disconnected from the

Strona 8

9 Declaration of Conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax: 00

Strona 9

GB KERN EMB 500-1GN Version 1.0 05/2014 Operating Instructions Precision balance Table of contents: 1 Technical data ...

Strona 10 - 6 Operation

1 Technical data KERN EMB 500-1GN Readout (d) 0,1 g Weighing range (Max) 500 g Reproducibility 0,1 g Linearity ± 0,2 g Adjusting weight (not inc

Strona 11

2 Fundamental information (general) 2.1 Intended use The balance you have acquired serves to determine the weighing value of the mate-rial to be we

Strona 12

3 Fundamental safety information 3.1 Observe the information in the operating instructions Carefully read this operation manual before setup and c

Strona 13

5 Unpacking, installation and commissioning 5.1 Place of installation, place of use The balance is constructed in such a way that reliable weighing

Strona 14

5.2 Unpacking Carefully remove the balance from its packaging, remove the plastic wrapping and position the balance in its intended working location.

Strona 15 - 8 Troubleshooting

5.6 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated – in compliance

Strona 16 - Declaration of Conformity

1 2 5.7 Underfloor weighing Objects which, because of their size or shape, cannot be put on the scale, can be weighed by means of underfloor weighing