KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-
FCB-BA-e-0911 10 4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual befor
FCB-BA-e-0911 11Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or r
FCB-BA-e-0911 12 6.4 Operation using a (rechargeable) battery (optional) Lift-off the battery cover on the lower side of the balance. Connect 9 V com
FCB-BA-e-0911 136.6 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating
7 Operation elements 7.1 Overview of display 7.2 Keyboard overview Key Designation Function PRINT-key • Calculate weighing data via interface
8 Basic Operation Start-up ONOFF Ö Press ON button. The balance will carry out a self-test The balance is ready for weighing when the weight display
Ö Weigh the material, the net weight will be indicated. 47.2g The taring process can be repeated any number of times, e.g. when adding several comp
Plus/minus weighings TARE For example unit weight control, fabrication control etc. Ö Put the nominal weight on the weighing plate and tare using th
• Back to weighing mode by pressing the MODE button. • Error message „Er 1“ Piece below minimum weight of piece (See chpt. 1 “Technical specificat
Percent determination MODECAL SETM Percentage calculation facilitates weight display in percent related to a reference weight equivalent to 100 %.
GB KERN FCB Version 1.1 01/2009 Operating Manual Compact balance Table of Contents 1 Technical Data ...
9 Menu 9.1 Navigation in the menu Access to menu PRINT 0.0g VnT In weighing mode keep the PRINT key pressed until [Unit] appears. How to select men
2. Change settings of several menu items SETM MODECAL PR RE CR Pl PC TARE ETmij SETM STQRE Acknowledge selected menu item using SET k
Save/reject and exit the menu SETM ETmij STQRE Ö Save SETM 0.0g Ö Reject PRINT 0.0g Either Any changes carried out are stored by pressing
FCB-BA-e-0911 239.2 Menu overview Description of function Function Parameter Description of options g* Gram kg Kilogram (dependent on model) oz Po
FCB-BA-e-0911 24 19200 9600* 4800 2400 Baud rate (see chapter 9.4). bAUd 1200 on* Automatic switch-off function after 3 min without changing lo
9.3 Description of individual menu items Weighing Units Ö In weighing mode keep the PRINT key pressed until [Unit] appears. 0.0g VnT Ö P
Dosing and Zero-tracking 0.0g VnT Tl The Auto-Zero function is used to tare small variations in weight automatically. In the event that small
FCB 3K0.1 FCB 6K0.5 FCB 6K1 FCB 8K0.1 1000 2000 2000 2000 2000 4000 4000 5000 3000 6000 6000 8000 FCB 12K1 FCB 15K5 FCB 16K0.2 FCB 24K1
Animal weighing function 0.0g VnT ANL The animal weighing function can be applied for busy weighing. During a defined period the average value of
Reset to factory setting 0.0g VnT RST This function resets all balance settings to factory setting. Ö In weighing mode keep the PRINT key presse
FCB-BA-e-0911 3 8 Basic Operation ... 15 Start-up...
9.4 Interface parameters Data output is carried out via interface RS 232 C. General The previous condition for the data transfer between balance an
printout 0.0g VnT LAPR Using this function data are selected which are to be sent via the RS232C (not valid for data transfer mode BAPr ). Ö In
Baud rate 0.0g VnT BAUD The baud rate defines the transfer speed vie the interface, 1 Baud = 1 Bit/second. Ö In weighing mode keep the PRINT key
FCB-BA-e-0911 3310.3 Explanation of the data transfer Pr PC: Press the PRINT key, at stable weight the format is transferred from LAPR. a. Forma
rE Cr: The remote control commands s/w/t are sent from the remote control unit to the balance as ASCII code. After the balance having received the s
FCB-BA-e-0911 3510.4 Output on bar code printer The data transfer mode has to be set on „BA Pr“ (chapter 9.4). As bar code printer the Zebra printer
FCB-BA-e-0911 36 12 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weig
FCB-BA-e-0911 4 1 Technical Data KERN FCB 3K0.1 FCB 6K0.5 FCB 6K1 Readability (d) 0.1 g 0.5 g 1 g Weighing range (max) 3 kg 6 kg 6 kg Tar
FCB-BA-e-0911 5 KERN FCB 8K0.1 FCB 12K1 FCB 15K5 Readability (d) 0.1 g 1 g 5 g Weighing range (max) 8 kg 12 kg 15 kg Taring range (subtractiv
FCB-BA-e-0911 6 KERN FCB 16K0.2 FCB 24K1 FCB 24K2 Readability (d) 0.2 g 1 g 2 g Weighing range (max) 16 kg 24 kg 24 kg Taring range (subtrac
FCB-BA-e-0911 7 KERN FCB 24K10 FCB 30K0.5 FCB 30K1 Readability (d) 10 g 0.5 g 1 g Weighing range (max) 24 kg 30 kg 30 kg Taring range (subtra
2 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postbox 4052 email: [email protected] Phone: 0049-[0]7433- 9933-0 Fax: 004
FCB-BA-e-0911 93 Basic Information (General) 3.1 Proper use The balance you purchased is intended to determine the weighing value of material to be
Komentarze do niniejszej Instrukcji