Kern-sohn FXN 30K-2 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Waga Kern-sohn FXN 30K-2. KERN & SOHN FXN 30K-2 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 22
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating manual
Compact balance
KERN FXN
Version 1.0
11/2014
GB
FXN-N-BA-e-1410
Przeglądanie stron 0
1 2 3 4 5 6 ... 21 22

Podsumowanie treści

Strona 1 - Compact balance

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-

Strona 2 - KERN FXN-N

6.2 Unpacking and placing Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation. Leve

Strona 3 - 1 Technical data

6.2.1 Scope of delivery Serial accessories:  Balance  Weighing pan  Operating manual  4 x 1.5V size D batteries 6.3 Battery operation The

Strona 4 - 1.1 Dimensions

 Insert 4 x 1.5V size D batteries.  Close the battery compartment cover.  Move both levers 90° to the right Ensure that the cables are no

Strona 5 - 2 Appliance overview

6.4 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (s

Strona 6 - 2.2 Keyboard overview

7.1.1 Adjustment Operation  Start balance by pressing .  Press and at the same time in weighing mode.  01 FnC appears  Select 02

Strona 7 - 3.3 Warranty

8 Operation 8.1 Weighing  Start balance by pressing .  The balance will carry out a self-test  As soon as the weight display shows „0.0“ an

Strona 8 - 5 Transport and storage

0.0 The tare weight is saved until it is deleted. Remove the load from the balance and press . The zero display appears and the triangle above the

Strona 9

9 The menu 9.1 Navigation in the menu  Start balance by pressing  In weighing mode press the and the button at the same time.  01 FnC

Strona 10 - 6.2 Unpacking and placing

9.2 Menu overview Menu item Function 01 FnC Access to the settings and functions of the balance. FnC 01 Setting of background lighting FnC 02

Strona 11 - 6.3 Battery operation

9.3 Set auto-OFF function  Start balance by pressing  Press and at the same time in weighing mode.  01 FnC appears  Confirm with

Strona 12

GB KERN FXN-N Version 1.0 11/2014 Operating manual Compact balance Contents 1 Technical data ...

Strona 13 - 7 Adjustment

10 Error messages Display Description Remedy E1 Zero range exceeded Unload the balance E2 Resetting range has not been reached. Check if the co

Strona 14

11 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must

Strona 15 - 8 Operation

12 Servicing, maintenance, disposal 12.1 Cleaning Before cleaning, please remove batteries from the appliance Please do not use aggressive cleanin

Strona 16 - 8.3 Background illumination

1 Technical data KERN FXN 3K -3 FXN 6K -3 FXN 10K -3 FXN 30K -2 Weighing range (max) 3 kg 6 kg 15 kg 30 kg Readability (d) 1 g 2 g 5 g 10

Strona 17 - 9 The menu

1.1 Dimensions 4 FXN-N-BA-e-1410

Strona 18 - 9.2 Menu overview

2 Appliance overview 2.1 Overview of display 2.1.1 Not verifiable models Display Designation Description Zeroing display Should the ba

Strona 19 - 9.3 Set auto-OFF function

2.2 Keyboard overview Button Designation Function ON-button, zero button Switch-on, set to zero OFF-button Switching Off TARE button Tare bal

Strona 20 - 10 Error messages

3 Basic Information (General) 3.1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed. It i

Strona 21 - 11 Instant help

4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and com