Kern-sohn GAB 30K-3PM Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Waga Kern-sohn GAB 30K-3PM. KERN & SOHN GAB 30K-3PM User Manual [de] [en] [fr] Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 39
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating manual
Table scales
KERN GAB-P
Version 1.1
03/2014
GB
GAB-P-BA-e-1411
Przeglądanie stron 0
1 2 3 4 5 6 ... 38 39

Podsumowanie treści

Strona 1 - Table scales

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-

Strona 2

Selection Function F In piece counting mode change between reference weight - item number - total weight UNIT In weighing mode: Weighing units swi

Strona 3

2.3 Dimensions 3 Basic Information (General) 3.1 Proper use The balance acquired by you serves to determine weight. It is intended to be used

Strona 4 - 1 Technical data

3.3 Warranty Warranty claims shall be voided in case • Our conditions in the operation manual are ignored • The appliance is used outside the d

Strona 5

4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and com

Strona 6 - 2 Appliance overview

6 Unpacking, Setup and Commissioning 6.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are a

Strona 7

6.2 Unpacking / installation Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.

Strona 8

6.3 Scope of delivery Serial accessories: • Balance, see chap. 2 • Mains adapter • Rechargeable battery • Operating manual 6.4 Mains connec

Strona 9

6.7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in complianc

Strona 10

6.8 Non-verifiable models:  Switch on the balance and during the selftest press the and the button at the same time. „UnLoad“ is displayed sh

Strona 11

 Carefully place adjusting weight in the centre of the weighing plate. Wait for stability display , then press . „PASS“ will be displayed shortly

Strona 12

GB KERN GAB-P Version 1.1 03/2014 Operating manual Table scale Contents 1 TECHNICAL DATA ...

Strona 13 - 5 Transport and storage

6.10.1 Position of seals and adjusting switch 1. Seal of approval 2. Cover 3. Housing screw 4. Adjustment switch 6.11 Checking the balance

Strona 14

7 Operation 7.1 Switching on/off  To turn on/off confirm forward at the underneath on the right of the scale and keep pressed for a little while

Strona 15

7.3 Taring  Deposit weighing receptacles (example)  After successful standstill control press the button. The zero display and „NET“ will app

Strona 16

7.4 Overload warning Overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoi

Strona 17

7.6 Percent weighing Percent weighing allows to display weight in percent, in relation to a reference weight.  Switch on balance, wait for zero

Strona 18

7.7 Parts counting Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain numb

Strona 19 - UUGeneral hints:

7.8 Weighing with tolerance range You can set an upper or lower limit when weighing with tolerance range and thus ensure that the material to be wei

Strona 20

Settings:  In weighing mode press , the left number flashes and the symbol „HI“ will be displayed.  Enter upper limit by pressing and .

Strona 21 - 7 Operation

7.9 Manual totalizing This function is used to add the individual weighing values to the summation memory and to provide a printout via the installe

Strona 22 - (example)

Display of the saved weighing data:  With the weighing platform unloaded, press and the number of weighing processes, followed by the total wei

Strona 23

7 OPERATION ... 21 7.1 Switching on/off ...

Strona 24

7.10 Automatic adding-up This function is used to add the individual weighing values automatically into the summation memory and to provide a printou

Strona 25

Display of the saved weighing data:  With the weighing platform unloaded, press and the number of weighing processes, followed by the total weig

Strona 26

8 The menu 8.1 Navigation in the menu: Call up menu Press and at the same time in weighing mode Menu block Select main menu Using the ind

Strona 27

8.2 Menu overview 8.2.1 Non-verifiable models: Menu block Main menu Menu item Submenu Available settings / explanation r dUAL not documented r

Strona 28

bP0* Acoustic signal bP 0 No audio sound during tolerance weighing bP 1 Audio sound when weight is within tolerance range bP 2 Audio sound when weigh

Strona 29

beeP 0* Acoustic signal beeP 0 No audio sound during tolerance weighing beeP 11 Audio sound when weight is within tolerance range beeP 2 Audio sound

Strona 30

9 Printer 9.1 Change paper roll  To open, lift green cover and pull upwards.  Insert new paper roll as shown on diagram.  Pul

Strona 31

10 Data output RS 232C The balance is typically equipped with a RS 232C interface. 10.1 Technical data  ASCII code  8 data bits  no parity bi

Strona 32 - 8 The menu

11 Maintenance, servicing, removal, error messages 11.1 Clean Before cleaning, disconnect the appliance from the operating voltage. Please do not us

Strona 33

12 Declaration of conformity KERN & Sohn GmbH D-72322 Balingen-Frommern Postbox 4052 E-Mail: [email protected] Phone: 0049-[0]7433- 9933-0 Fax

Strona 34

1 Technical data KERN GAB 6K-5P GAB 10K-4P GAB 30K-4P Weighing range (max) 6 kg 12 kg 30 kg Readability (d) 0.05 g 0.1 g 0.2 g Reproducibil

Strona 35

KERN GAB 6K-3PM GAB 10K-3PM GAB 30K-3PM Weighing range (max) 3 kg / 6 kg 6 kg / 15 kg 15 kg / 30 kg Readability (d) 1 g / 2 g 2 g / 5 g 5 g /

Strona 36 - 9 Printer

2 Appliance overview 1. Weighing pan 2. Printer 3. Buttons 4. Bubble level 5. RS 232 interface 6. Battery compartment 6 GAB-P-BA-e-1

Strona 37 - 10 Data output RS 232C

7. Transport guard screw 8. Levelling screw 9. Connection of mains adapter 10. ON/OFF switch 11. Seal of verifiable models 12. Seal with an und

Strona 38

2.1 Overview of display Display Description Description 1 The battery capacity display appears when the capacity of the battery is getting lo

Strona 39 - Declaration of conformity

1 2 2.2 Keyboard overview 1. Bubble level Level balance with foot screws until the air bubble of the water balance is in the presc

Komentarze do niniejszej Instrukcji

Brak uwag