Instruction Manual
FH 10– 500 EXT
FH_10-500-BA-e-1213 6
8. CE Declaration of conformity
FH Adjustment Procedure
instrument
light is on
adjustment
procedure
ON/OFF, press PEAK
and PRINT together for
more times and very
short intervals until the
red light is on.
red light is
on
Type
Directly after pressing
PEAK + PRINT
shown
a (Back to the
normal
mode ???)
(If you are in the
operation mode again,
the instrument has to be
shut off. Start again with
1. Eventually pressing
the buttons faster)
instrument
type
correspondi
ng value to
the instr.
appears in
the display
instrument
type
adjustment
sequence
red light is
on
adjustment
weight
with▼▲ specify the
calibration weight in
Newton (X kg * 9,81)
is written in
the display
adjustment
weight
simultaneously
9 Attach the
weight to the
instrument
(by hanging)
Press ZERO The
instrument
changes to
operation
mode.
Adjustment
procedure is
finished